首頁 TOS5200 Series

AC耐壓試驗儀:5kV
TOS5200 Series

輸出為5kV/100mA (AC), IEC, EN, UL, VDE, JIS等各國安全規定或電器產品安全法的要求之上, 可對電子機械, 電子部品進行耐電壓測試。
因為不受AC 輸入端電壓和頻率變化的影響, 可保持安定的輸出電壓, 所以在電源環境不安定的區域也可進行安定, 安心的測試。這是一款基於操作性 / 實用性 / 安全性 / 低價格為一體的AC 耐電壓測試儀。


  • 可以對施加電壓進行時間控制

    [搭載輸出電壓的上升時間 / 下降時間的控制功能!] 防止超規格測試對被測物產生不良的影響。

詳細規格

Withstand voltage tester section
AC Output section Output range   0.05 kV to 5.00 kV
Accuracy ± (2 % of set + 20 V) when no load is connected
Operating range 0.00 kV to 5.50 kV
Resolution 10 V STEP
Max. rated output *1 500 VA (5 kV/100 mA)
Max. rated voltage 5 kV
Max. rated current 100 mA (when the output voltage is 0.5 kV or greater)
Transformer rating 500 VA
Output voltage waveform *2   Sine
Distortion If the output voltage is 0.5 kV or more: 3 % or less (when no load or a pure resistive load is connected)
Crest factor √ 2 ± 3 % less than (when the output voltage is 800 V or greater, no load)
Frequency   50 Hz or 60 Hz
Accuracy ± 0.5 % (excluding during voltage rise time)
Voltage regulation 10 % or less (when changing from maximum rated load to no load)
Input voltage variation ±0.3 % (5 kV when no load is connected; power supply voltage: 90 V to 250 V)
Short-circuit current 200 mA or more (when the output voltage is 1.0 kV or greater)
Output method PWM switching
Start voltage The voltage at the start of withstanding voltage tests can be set to 50 % of the test voltage.
Limit voltage The test voltage upper limit can be set . AC: 0.00 kV to 5.50 kV
Output voltage monitor feature If output voltage exceeds the specified value + 350 V or is lower than the specified value – 350 V, output is turned off, and protective features are activated.
Voltmeter Digital Measurement range 0.000 kV to 6.500 kV AC
Display □ . □□□ kV
Accuracy V < 500 V:± (1.5 % of reading + 20 V)、V ≧ 500 V: ±1.5 % of reading
Response*3 True rms, Average value response/rms display switchable
Hold feature After a test is finished, the measured voltage is retained until the PASS or FAIL judgment is cleared.
Ammeter Digital Measurement range AC:0.00 mA to 110 mA
Display i = measured current
i < 1 mA:□ . □□□ mA
1 mA ≦ i < 10 mA:□ . □□□ mA
10mA ≦ i < 100 mA:□□ . □□ mA
100 mA ≦ i:□□□ . □ mA
Accuracy*4 1.00 mA ≤ i: ± (1.5 % of reading), i < 1.00 mA: ± (1.5 % of reading + 30 μA)
Response*3 True rms, Average value response/rms display switchable
Hold feature After a test is finished, the measured current value is retained until the PASS judgment is cleared.
 
Judgment feature Judgment method and judgment operation
Judgment Judgment method Display Buzzer SIGNAL I/O
UPPER FAIL If a current that is greater than or equal to the upper limit is detected, the output is turned off, and an UPPER FAIL judgment occurs. FAIL LED lights; UPPER is displayed on the screen ON Generates a U-FAIL signal
LOWER FAIL If a current that is less than or equal to the lower limit is detected, the output is turned off, and a LOWER FAIL judgment occurs. This judgment is not performed during voltage rise time (Rise Time) of all tests and during the voltage fall time (Fall Time) of AC withstanding voltage tests. FAIL LED lights; LOWER is displayed on the screen ON Generates a U-FAIL signal
PASS If the specified time elapses without any problems, the output is turned off, and a PASS judgment occurs. PASS LED lights
displayed on the screen
ON Generates a PASS signal
 
· If PASS HOLD is enabled, the PASS signal is generated continuously until the TOS5300 Series receives a STOP signal.
· The UPPER FAIL and LOWER FAIL signals are generated continuously until the TOS5300 Series receives a STOP signal.
· The FAIL and PASS buzzer volume levels can be changed.
· For PASS judgments, the length of time that the buzzer sounds for is fixed to 0.2 seconds. Even if PASS HOLD is enabled, the buzzer turns off after 0.2 seconds.
Upper limit setting AC: 0.01 mA to 110 mA
Lower limit setting AC: 0.01 mA to 110 mA / OFF
Judgment accuracy *4 1.00 mA ≤ i: ± (1.5 % of set), i < 1.00 mA: ± (1.5 % of set + 30 μA)
Current detection method Calculates the current’s true rms value and compares this value with the reference value
Calibration Calibrated with the rms of a sine wave using a pure resistive load
Time Voltage rise time   0.1 s to 10.0 s
Resolution 0.1 s
Voltage fall time 0.1 s / OFF (only enabled when a PASS judgment occurs)
Test Time   0.1 s to 999 s, can be turned off (TIMER OFF)
Resolution 0.1 s to 99.9 s: 0.1 s0 / 100 s to 999 s: 1 s
Accuracy ±(100 ppm + 20 ms) excluding Fall Time
Ambient temperature Upper limit setting Pause time Output time
t ≤ 40 °C AC 50 mA < i ≤ 110 mA Greater than or equal to the output time 30 min. max.
i ≤ 50 mA Not necessary Continuous output possible
Output voltage 1 kV 2 kV 5 kV
When using 350 mm long test leads that are suspended in air (TYP) 2 μA 4 μA 10 μA
When using the accessory, high test lead TL31-TOS (TYP) 16 μA 32 μA 80 μA
Other features/interfaces
Test mode Double action feature Tests can only be started by pressing and releasing STOP and then pressing START within 0.5 seconds of releasing the STOP switch.
Length of time to maintain a PASS judgment result You can set the length of time to maintain a PASS judgment: 50 ms, 100 ms, 200 ms, 1 s, 2 s,5 s, or HOLD.
Momentary feature Tests are only executed while the START switch is held down.
Fail mode feature This feature enables you to prevent remotely transmitted stop signals from clearing FAIL judgments and PROTECTION modes.
Timer feature This feature finishes tests when the specified time elapses.
Output voltage monitor feature If output voltage exceeds “setting + 350 V” or is lower than “setting – 350 V,” the TOS5200 switches to PROTECTION mode, output is turned off, and testing finishes.
Memory Up to three sets of test conditions can be saved to memory.
Key lock Locks panel key operations (settings and changes).
Protection functions   Under any of the following conditions, the TOS5200 switches to the PROTECTION state, immediately turns output off, and stops testing. A message is displayed on the screen.
Interlock Protection An interlock signal has been detected.
Power Supply Protection An error was detected in the power supply.
Volt Error Protection While monitoring the output voltage, a voltage outside of the rated limits was detected. AC or DC withstanding voltage tests: ±350 V
Over Load Protection During a withstanding voltage test, a value that is greater than or equal to the output limit power was specified. AC withstanding voltage test: 550 VA.
Over Heat Protection The internal temperature of the TOS5200 became too high.
Over Rating Protection During a withstanding voltage test, the output current was generated for a length of time that exceeds the regulated time.
Remote Protection A connection to or disconnection from the front- panel REMOTE connector was detected.
SIGNAL I/O Protection The rear-panel SIGNAL I/O connector’s ENABLE signal has changed.
USB Protection The USB connector has been disconnected while the TOS5200 was being controlled through the USB interface.
Interface USB USB Specification 2.0
RS232C*1 D-SUB 9-pin connector on the rear panel (compliant with EIA-232-D) All functions other than the POWER switch and KEY-LOCK
REMOTE Front-panel 9-pin MINI DIN connector. By connecting an optional device to this connector, you can control the starting and stopping of tests remotely.
SIGNAL I/O Rear-panel D-sub 25-pin connector
Talk mode Description
0 It responds only for commands from PC. (Default setting)
1 It responds automatically for start and end test, and returns the status, setting value, measured value.
Response at start
Response at end of test Status ,,,,
Setting value, Measured value Test No., Programme No., Test mode, Measured voltage, Measured current, Test time

SIGNAL I/O connector pinout

PIN No Signal name I/O Rear-panel D-sub 25-pin connector
1 INTERLOCK+ I If you open the positive and negative terminals, the output is turned off, and the TOS5300 Series is switched to Protection mode.
Open: The resistance between the two terminals is 1.2 kΩ or greater. Short: The resistance between the two terminals is 1 kΩ or less.
2 PM0 I
PM0 PM1 Called Panel Memory Number
H H Memory 1
L H Memory 2
H L Memory 3
L L –
 
Panel memory selection signal. The selection signal is latched on the rising edge of the input strobe signal to recall panel memory.
* The selection of memory is prioritized over TEST SEL and AUTO SEL.
3 PM1 I
4 NC − −
5 NC − −
6 NC − −
7 NC − −
8 NC − −
9 STB I Panel memory strobe signal input terminal.
10 TEST SEL I NA
11 AUTO SEL I NA
12 COM − Circuit’s common terminal
13 INTERLOCK- I If you open the positive and negative terminals, the output is turned off, and the TOS5300 Series is switched to Protection mode.
Open: The resistance between the two terminals is 1.2 kΩ or greater. Short: The resistance between the two terminals is 1 kΩ or less.
14 HV.ON O ON during test and while voltage remains between output terminals
15 TEST O ON during test (except when voltage is rising or falling)
16 PASS O On for at least 0.2 seconds (the PASS HOLD time) when a PASS judgment occurs. On continuously when the PASS HOLD time is set to HOLD.
17 U-FAIL O Continuously ON when value over acceptable maximum is detected, and UPPER FAIL is determined
18 L-FAIL O Continuously ON when value under acceptable minimum is detected, and LOWER FAIL is determined
19 READY O ON during standby (READY status)
20 PROTECTION O ON while protection function is activated (PROTECTION ON)
21 START I Start signal input terminal
22 STOP I Stop signal input terminal
23 ENABLE I Start signal’s ENABLE signal input terminal: Shifts to the Protection status when the ENABLE signal changes
24 +24V − +24V internal power supply output terminal: Maximum output current 100 mA
25 COM − Circuit’s common terminal
General Specifications
Display LCD: LED backlight
Environment Installation location Indoors, at a height of up to 2000 m
Spec guaranteed range temperature/ humidity 5 °C to 35 °C (41 °F to 95 °F)/20 %rh to 80 %rh (no condensation)
Operating range temperature/ humidity 0 °C to 40 °C (32 °F to 104 °F)/20 %rh to 80 %rh (no condensation)
Storage range temperature/ humidity -20 °C to 70 °C (-4 °F to 158 °F)/90 %rh or less (no condensation)
Power supply Nominal voltage range (allowable voltage range) 100 VAC to 240 VAC (90 VAC to 250 VAC)
Power consumption When no load is connected (READY) 100 VA or less
When rated load isconnected 800 VA max.
Allowable frequency range 47 Hz to 63 Hz
Insulation resistance (between AC LINE and the chassis) 30 MΩ or more (500 VDC)
Withstanding voltage (between AC LINE and the chassis) 1500 VAC, one minute
Earth continuity 25 AAC, 0.1 Ω or less
Electromagnetic compatibility (EMC) *1 Complies with the requirements of the following directive and standard.
EMC Directive 2014/30/EU
EN 61326-1(ClassA *2), EN 55011(ClassA *2, Group1 *3)
EN 61000-3-2, EN 61000-3-3
Applicable under the following conditions
The maximum length of all cabling and wiring connected to the TOS5200 must be less than 2.5 m.
The shielded cable is being used when using the SIGNAL I/O.
The high test lead TL31-TOS
Safety *1 Complies with the requirements of the following directive and standard.
Low Voltage Directive 2014/35/EU, EN 61010-1 (Class I *4 Pollution degree 2)
Dimensions (mm(inches))(maximum) 320 (12.6″) (330(12.99″)) W × 132(5.2″) (150(5.91″)) H × 350(13.78″) (420(16.54″)) D
Weight Approx. 14 kg (30.9 lbs)
Accessories Power cord : 1pc. / High test lead (TL31-TOS) : 1set (1 red wire and 1 black wire, each with alligator clips); 1.5 m / D-sub 25-pin plug : 1set ; assembly type / High-voltage warning sticker : 1pc. / Setup Guide / Quick Reference(1 each for English and Japanese) / Safety informaion / CD-R *5