接地導通試驗儀:60A 地線電阻測量:30A
TOS6210
符合標準試驗的接地導通測試儀
TOS6210 本機承襲傳統產品(TOS6200)的基本性能與各項功能,例如採用恆電流驅動方式,可獲得失真小的電流波形,具備高測量精度;同時將最大試驗電流由 30A 擴大至標準所要求的 60A。此外,亦可針對標準規範之電壓降進行判定。
主機面板內建記憶體可預先儲存資訊處理設備、家電、醫療設備、測量儀器等 20 種安全標準的試驗條件,僅需簡單呼叫即可設定以 UL60950-1 為代表,包含 IEC、JIS 等規範之**保護接地(Protective Earthing)與保護連接(Protective Bonding)**導通試驗。
此外,亦配備多項滿足現場細部需求的功能,例如偏置消除功能、可輸入校正年月日與製造編號,並可透過 GPIB/RS-232C 讀出並記錄相關資料。
-

-
產品特性
採用恆定電流方式,測試電流:6~62A/阻抗:0.001~0.600Ω 利用視窗比較方式進行判定 可控制利用GPIB、RS-232C等外部傳輸電流、時間、電阻值 可讀回GPIB、RS-232C·測試值、試驗結果 可搭載100個記憶體功能和程式功能 設有接觸校驗功能 設有容易準確判斷狀態的大型顯示幕 可判定電阻值·電壓下降
詳細規格
Output
| Current setting range *1 *4 | Resolution 6.0 to 62.0 A AC (With respect to resistance resulting in output power of the maximum rated Output or less and an output terminal voltage of 5.4 V or less) | |
| Resolution | 0.1 A | |
| Accuracy | ± (1% of setting + 0.4A) | |
| Maximum rated output | 220 VA (at the output terminals) | |
| Distortion Rate | 2% or less (with respect to 0.1 Ω pure resistance load of 20 A or greater) | |
| Frequency | 50/60 Hz, sine wave (selectable) | |
| Accuracy | ±200 ppm | |
| Open terminal voltage | 6 Vrms or less | |
| Output method | PWM switching | |
Output ammeter
| Measurement range | AC 0.0A to 66.0A |
| Resolution / Accuracy | 0.1A / ± (1% of reading + 0.4A) |
| Response | Mean value response/rms value display (response time: 200 ms) |
| Hold function | The current measured at the end of test is held during the PASS or FAIL inteval |
Output voltmeter
| Measurement range | 0.00 to 6.00 V AC |
| Resolution / Accuracy | 0.01V / ± (1% of reading + 0.02V) |
| Offset cancel function | 0.00 to 5.40 V (Offset ON/OFF function provided) |
| Response | Mean value response/rms value display (response time: 200 ms) |
| Hold function | The voltage measured at the end of test is held during the PASS or FAIL inteval |
Ohmmeter *2
| Measurement range | 0.001 to 0.600 Ω |
| Resolution / Accuracy | 0.001 Ω / ± (2% of reading + 0.003 Ω) |
| Offset cancel function | 0.000 to 0.600 Ω (Offset ON/OFF function provided) |
| Hold function | The resistance measured at the end of test is held during the PASS or FAIL interval |
Pass/fail judgement function *3
| Resistance value-based judgement | Window comparator system · If a resistance value equal to or greater than the upper reference value is detected, a FAIL determination is returned. · If a resistance value equal to or less than the lower reference value is detected, a FAIL determination is returned. · If a resistance value has been judged as FAIL, the tester shuts off the output and generates a FAIL signal. · If the set time elapses without abnormalities, the tester shuts off the output and generates a PASS signal. |
|
| Setting range for the upper reference value (UPPER) | 0.001 to 0.600 Ω | |
| Setting range for the lower reference value (LOWER) | 0.001 to 0.600 Ω | |
| Resolution | 0.001Ω | |
| Judgment accuracy | ± (2% of UPPER + 0.003 Ω) | |
| Sampled voltage value-based judgment | Window comparator system · If a voltage value equal to or greater than the upper reference value is detected, a FAIL determination is returned. · If a voltage value equal to or less than the lower reference value is detected, a FAIL determination is returned. · If a resistance value has been judged as FAIL, the tester shuts off the output and generates a FAIL signal. · If the set time elapses without abnormalities, the tester shuts off the output and generates a PASS signal. |
|
| Setting range for the upper reference value (UPPER) *4 | 0.01 to 5.40 V | |
| Setting range for the lower reference value (LOWER) | 0.01 to 5.40 V | |
| Resolution | 0.01V | |
| Judgment accuracy | ± (2% of setting + 0.05 V) | |
| Calibration | Calibration is performed with the rms value of the sine wave, using a pure resistance load. | |
| LED | PASS | Lights for approximately 0.2 sec when the measured value has been judged as PASS. It is lit continuously when the PASS holding time is set to HOLD. |
| UPPER FAIL | Lights if a resistance or voltage value equal to or greater than the upper reference value is detected and judged FAIL. | |
| LOWER FAIL | Lights if a resistance or voltage value equal to or greater than the upper reference value is detected and judged FAIL. | |
| Buzzer | The buzzer sounds for the pass holding time has been set if the measured value has been judged as PASS. · The buzzer sounds continuously under the following condition: The measured value has been judged as PASS when the PASS holding time is set to HOLD. The measured value has been judged as UPPER FAIL. The measured value has been judged as LOWER FAIL. · The buzzer volume for FAIL or PASS judgment are adjustable. Note that it cannot be adjusted individually since setting is shared with the setting for PASS. |
|
Power supply
| Input voltage range | 85 to 250 V AC | |
| Power consumption | At no load (READY) | 60 VA or less |
| Rated load | 420 VA max. | |
| Input frequency range | 47 Hz to 63 Hz | |
| Insulation resistance | 30MW min. (500 Vdc), between AC line and chassis | |
| Withstanding voltage | 1390 Vac (2 seconds), between AC line and chassis | |
| Ground bond | 25 Aac/ 0.1 Ω max | |
*實際規格依型號與選配項目而異,詳情請洽啟宸科技。