洩漏電流測試器
TOS3200

TOS3200是一款可適用於實施除醫用電氣設備以外的常規電氣設備漏電流 (接觸電流、保護導體電流) 試驗儀。可實施符合IEC、UL、JIS、電氣產品安全法等標準要求的試驗。主機內置的記憶體儲存了51種廣及資訊技術、家電、Audio/Video、照明、電動工具、測量和控制設備儀器有關的IEC/JIS標準、以及電氣產品安全法中規定的試驗條件, 從而能夠便於利用面板操作, 實施符合標準要求的試驗。


  • 產品特性

    測量3種操作模式的洩漏電流:觸控電流(TC)、保護導體電流(PCC)和METER操作模式 測試儀中包含八種測量電路網路,用於測量一般電器的觸控電流 可以測量DC/RMS/PEAK,實現以30 mA為最大RMS的「真均方根值」 易於理解的操作 可以在獨立測試和自動測試之間進行選擇 51種預設安全標準測試 控制校準過期限制的功能
  • 規格

    DC/RMS︰30μA~30mA、PEAK︰50μA~90mA

詳細規格

Specifications

Measurement item, measurement mode

Measurement items 3 types, namely, touch current (TC) measurement, protective conductor current (PCC) measurement, and METER
Measurement method TC Measure the voltage drop across the reference resistor, using a measurement circuit network (NTWK), and then calculate the current.
PCC Measure the voltage drop across the reference resistor connected to the protective earth wire, and then calculate the current.
METER Measure the voltage and current using the measurement terminals.
Measurement mode DC/RMS/PEAK (RMS being the true root-mean-square value)
Measurement circuit network (NTWK) Network A Basic measurement element: (1.5 kΩ//0.22 μF) + 500 Ω
Network B/B1 Basic measurement element: (1.5 kΩ//0.22 μF) + 500 Ω//(10 kW + 0.022 μF)
Network C Basic measurement element: (1.5 kΩ//0.22 μF) + 500 Ω//(10 kΩ + (20 kΩ + 6.2 nF)//9.1 nF)
Network D Basic measurement element: 1 kΩ
Network E Basic measurement element: 1 kΩ//(10 kΩ + 11.225 nF + 579 Ω)
Network F Basic measurement element: 1.5 kΩ//0.15 μF
Network G Basic measurement element: 2 kΩ
Network constant tolerance Resistance: ±0.1%, capacitor 0.15 μF: ±2%, other: ±1%

Current measurement section

Measurement range Range 1 DC/RMS: 30 μA to 600 μA, PEAK: 50 μA to 850 μA *3
Range 2 DC/RMS: 125 μA to 6.00 mA, PEAK: 175 μA to 8.50 mA *3
Range 3 DC/RMS: 1.25 mA to 30.0 mA, PEAK: 1.75 mA to 90.0 mA *3
Range switching AUTO/FIX
Measured current (i) display/resolution i < 1mA: □□□μA/1 μA, 1 mA i < 10 mA: □.□□mA/0.01 mA 10 mA i < 100 mA: □□,□mA/0.1 mA
Measurement accuracy *5 Range 1 DC ±(5.0% of rdng + 20 μA)
RMS 15Hz≦f≦10kHz:±(2.0% of reading+8μA)
10kHz<f≦1MHz:±(5.0% of reading+10μA)
PEAK 15Hz≦f≦10kHz:±(5.0% of reading+10μA)
Range 2 DC ±(5.0% of reading+50μA)
RMS 15Hz≦f≦10kHz:±(2.0% of reading+20μA)
10kHz<f≦1MHz:±(5.0% of reading+20μA)
PEAK 15Hz≦f≦1kHz:±(2.0% of reading+50μA)
1kHz<f≦10kHz:±(5.0% of reading+50μA)
Range 3 DC ±(5.0% of reading+0.5mA)
RMS 15Hz≦f≦10kHz:±(2.0% of reading+0.2mA)
10kHz<f≦1MHz:±(5.0% of reading+0.2mA)
PEAK 15Hz≦f≦1kHz:±(2.0% of reading+0.5mA)
1kHz<f≦10kHz:±(5.0% of reading+0.5mA)
Input resistance, input capacitance 1 MΩ±1 %, <200 pF
Common mode rejection ratio f ≦ 10 kHz: 60 dB or greater, 10 kHz < f ≦ 1 MHz: 40 dB or greater

Judgment function

Judgment method Pass/fail judgement by setting upper and lower current limits in window comparator mode
Judgment U-FAIL for currents above the upper limit; L-FAIL for currents below the lower limit.
Display U-FAIL/L-FAIL/PASS display, buzzer sounding
PASS hold The time for which a PASS judgement is retained can be set to 0.2 s to 10.0 s or to HOLD
Setting range Range 1 DC/RMS: 30 μA to 600 μA, PEAK: 50 μA to 850 μA *4
Range 2 DC/RMS: 151 μA to 6.00 mA, PEAK: 213 μA to 8.50 mA *4
Range 3 DC/RMS: 1.51 mA to 30.0 mA, PEAK: 2.13 mA to 90.0 mA *4
Judgment accuracy Conforms to measurement accuracy. (Read rdng as set.)

Measurement of voltage between A and B

Measurement range DC/RMS: 10.000 V to 300.0 V, PEAK: 15.000 V to 430.0 V
Accuracy ±(3% of rdng + 2 V), measurement range fixed at AUTO
Input impedance Approx. 40 MΩ
SELV detection Set the SELV to detect; if this value is exceeded, the DANGER lamp is turned ON
SELV setting range 10 V to 99 V, in 1-V steps, OFF function provided

Timer, test execution function, memory

Timer Test wait time Setting range: 0 s to 999 s, accuracy: ±(100 ppm of set + 20 ms)
Test time Setting range: 1 s to 999 s/OFF function, accuracy: ±(100 ppm of set + 20 ms)
Test execution Auto test (AUTO): Automatic execution of up to 100 steps (test conditions)
Independent test (MANUAL): Independent execution of TC, PCC, or METER measurement
memory Test conditions AUTO: Up to 100 sequence programs can be saved (up to 500 steps in total).
MANUAL: Up to 100 sequence programs can be saved.
Test results The user can select whether to save the judgement results when they are output at the end of the tests.
AUTO: Test results for up to 50 programs can be recorded.
MANUAL: Test results for up to 50 tests can be recorded.

Other functions

Measured value conversion (CONV) Converts the measured current value into the corresponding value at the preset power voltage
Setting range: 80.0 V to 300.0 V, OFF function provided
MEASURE MODE Selects a measured value from those below
NORM: Displays the measured value in the measurement period
MAX: Displays the largest measured value in the measurement period
Power positive/negative phase selection (POL) NORM: Positive phase connection, REVS: Negative phase connection
Single fault selection (COND) NORM: Normal, FLTNEU: Disconnection of the neutral wire, FLTPE: Disconnection of the protective earth wire
Ground check Generates CONTACTFAIL if the enclosure is grounded in a TC (EncLiv, EncNeu) test
MEASURE CHECK Checks the measurement function between measurement terminals A and B, and places the tester in the PROTECTION state if an error is detected
Voltage measurement(EUT) Measurement range: 80.0 V to 250.0 V, resolution: 0.1 V, accuracy: ±(3% of rdng + 1 V)
Current measurement(EUT) Measurement range: 0.1 A to 15.00 A, resolution: 0.01 A, accuracy: ±(5% of rdng + 30 mA)
Power measurement (effective power) Measurement range: 10 W to 1500 W
Accuracy (at a power voltage of 80 V or higher and a load power factor of 1): ±(5% of rdng + 8 W)
System clock Recording Items: Calibration date and time, test date and time, permissible date and time: Up to 2099
Calibration time limit management(CAL PROTECT) Enables the setting of a calibration time limit. Once this time has passed, a warning is output at power on
ON: Places the tester in the PROTECTION state (disables the use of the tester), OFF: Displays warning.
Protective operation Relay operation error, overload, over range, measurement function check, failure of internal battery, etc.

Interface

RS232C D-Sub 9-pin connector (conforming to EIA-232D), baud rate: 9600/19200/ 38400 bps (For connection to a PC, use a “9-pin female-female reverse” cable.)
GPIB Conforms to IEEE Std. 488-1978. (SH1,AH1,T6,TE0,L4,LE0,SR1,PP0,DC1,DT0,C0,E1)
USB USB Specification2.0
REMOTE 6-pin MINIDIN connector (for HP21-TOS (separately sold option) only)
SIGNAL I/O 25-pin D-Sub connector

General

Measurement terminal Rated voltage/current Terminals A to B: 250 V, terminal to chassis: 250 V, 100 mA
Measurement category CATⅡ
Valid terminal display Terminals effective to measurement are indicated with LED lamps.
Environment Warranty range Temperature: 5 ̊C to 35 ̊C, humidity: 20% rh to 80% rh (no condensation)
Operating range Temperature: 0 ̊C to 40 ̊C, humidity: 20% rh to 80% rh (no condensation)
Storage range Temperature: –20 ̊C to 70 ̊C, humidity: 90% rh or less (no condensation)
Installation location Indoors, altitude of 2000 m or less
AC line power requirements Input power Nominal input rating:100Vac to 240Vac, 50/60Hz, power consumption: 70 VA max.
for EUT Nominal input rating:100Vac to 240Vac, 50/60Hz
Rated output capacity: 1500 VA, maximum current: 15 A, rush current: 70 A peak max. (within 20 ms)
Insulation resistance 30 MΩ or greater (500 Vdc) (between AC line and chassis, between measurement terminal and chassis)
Withstanding voltage 1390 Vac, 2 seconds/20 mA or less (between AC line and chassis)
Earth continuity 25 Aac/0.1 Ω or less
Safety Conforms to the requirements of the directive and standard below.
Low Voltage Directive 2014/35/EU, EN61010-1 (Class I, Pollution degree 2)
Electromagnetic compatibility (EMC)*1*2 Complies with the requirements of the following directive and standards.
EMC Directive 2014/30/EU, EN 61326-1 (Class A), EN 55011 (Class A, Group 1), EN61000-3-2, EN61000-3-3
Applicable conditions: All cables and wires used to connect to this product must be shorter than 3 meters.
Use the supplied test leads.
Outside dimensions, weight 320[12.60 inch] (345[13.58 inch]) W × 88[3.46 inch] (105[4.13 inch]) H × 270[10.63 inch] (335[13.19 inch]) D mm, approx. 5 kg(approx. 11.02 lbs)
Accessories 1 set of test leads (TL21-TOS: red and black, one each, with alligator clips) 1 flat probe (FP01-TOS), 1 spare fuse (15 A, for EUT power) 1 instruction manual, 1 circuit principle diagram sticker 2 power cords (for the tester and for the EUT AC line)
 
*實際規格依型號與選配項目而異,詳情請洽啟宸科技。